IR/UV Source Stimulator >
IR Stimulator Spec Sheet
Dimensions
and Weight
| Width
x Depth x Height: |
15
¾” x 12 ¾” x 8” |
| Weight: |
35
lbs |
Source
| Description: |
Cylindrical
ceramic core heated by platinum resistance wire |
| Nominal
Radiating Area: |
3 mm (diameter) x 10 mm (length) |
| Nominal
Temperature: |
1450o K |
| Spectral
Content: |
1 – 25 mm |
| Radiant
Intensity: |
5.9
W/sr (2.5 – 5 mm)
3.8 W/sr
(8 – 14 mm) |
Source
Modulation
| Description: |
Linear
variable neutral density filter positioned with computer controlled
motion stage; absolute levels can be set with neutral density filters
placed in the beam path (inside the stimulator housing) |
| Dynamic
Range: |
3 OD (optical densities) |
| Modulation
Rate: |
> 3 OD / sec |
| Step
Size: |
< 2 x 10-4 OD |
Shutter
System
| Description: |
Dual
overlapping blade electronic shutter under computer control |
| Aperture
Diameter: |
6
mm |
| Transition
Time: |
1
ms |
| Repetition
Rate: |
Variable
up to 200 Hz |
Output
| Description: |
Output
beam is collimated with a collimating lens external to the stimulator
housing; the output beam diameter is set with an aperture stop placed
after the collimating lens |
| Measured
Output Radiant Intensity: |
1.77
W/sr in 2.5 – 5 mm band (max)
0.75
W/sr in 8 – 14 mm band (max) |