IR/UV Source Stimulator >
UV Stimulator Spec Sheet
Dimensions
and Weight
| Width
x Depth x Height: |
15
¾” x 12 ¾” x 8” |
| Weight: |
35
lbs |
Source
| Description: |
Low
pressure deuterium arc lamp (ozone free) |
| Arc
Diameter: |
0.5 mm |
| Spectral
Content: |
180 – 400 nm |
| Radiant
Intensity: |
1.0 x 10-4 W/sr-nm @ 250 nm |
Source
Modulation
| Description: |
Linear variable neutral
density filter positioned with computer controlled motion stage;
absolute levels can be set with neutral density filters placed in
the beam path (inside the stimulator housing) |
| Dynamic
Range: |
> 4 OD (optical densities) |
| Modulation
Rate: |
> 4 OD / sec |
| Step
Size: |
< 2 x 10-4 OD |
Shutter
System
| Description: |
Dual overlapping blade electronic
shutter under computer control |
| Aperture
Diameter: |
6
mm |
| Transition
Time: |
1
ms |
| Repetition
Rate: |
Variable
up to 200 Hz |
Output
| Description: |
Output
beam is collimated with a collimating lens external to the stimulator
housing; the output beam diameter is set with an aperture stop placed
after the collimating lens |
| Measured
Output Radiant Intensity: |
1.3
x 10-4 W/sr-nm @ 265 nm (max) |